Dongguan julong signing our with rock wool insulation heat load test device - order no. 20120313001
KP8096 with rock wool insulation heat load test device order production requirements:
A, scope of application:
L suitable for rock wool, mineral wool and glass cotton and products. Under fixed load, to a certain heating rate heating sample, meet the thickness reduction, through calculation, using interpolation method to find the thermal shrinkage temperature under load.
Second, conform to the standard:
In line with the GB/T11835 standards.
Three, technical parameters
Heat load test unit consists of a heating furnace, heating container and the temperature sensor.
L environment temperature: room temperature ~ 40 ℃ relative humidity: 75% or less
L working voltage: ac 220 v + / - 10 v, 50 hz
L heating power: 1000 w
L scale measuring range: at room temperature to 900 ℃
L heating rate: 5 ℃ / min plus or minus 0.3 ℃ and 3 ℃ / min plus or minus 0.2 ℃ can be set
L instrument using single chip computer, multipoint temperature programmed.
L heat source USES the hot plate, safe and reliable, long service life
L instrument equipped with surface format printers, can timely print test results.